SERVICE | general information

Colormetry CMU 324 series devices are low maintenace systems

  • the device does not need to be cleaned | → Please protect the device from external contamination
  • there are no spare parts to service or exchange
  • it gives the reagent cartridge as a consumable 
        → When the cartridge is replaced, all wearing parts in the interior of the device are automatically replaced
        [Consumption approx. 2-4 cartridges per year depending on the parameter settings]
  • There is the fine filter insert as a wearing part
        The service life of the fine filter depends on the quality of the feed water and can not be precisely defined
        [Exchange as needed | if no alarm is displayed, the manufacturer's recommendation is: change filter once a year]